4.13 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

102

■■■■■■■■

850 Professional IC –Anion – MCS – Prep 3

4.13.3.2

Cleaning the MSM

It may be necessary to clean the MSM in the following cases:

Increased backpressure onto the output capillaries of the MSM.

Blockage of the MSM which cannot be eliminated (solutions can no
longer be pumped through the MSM).

Jamming of the MSM which cannot be eliminated (MSM can no longer
be switched over).

1

2

3

4

5

Figure 54

MSM – Components

1

Union nut

2

MSM connecting piece 6.2832.010

3

MSM rotor A 6.2832.000

4

MSM housing

5

Slot in the MSM housing

Cleaning the MSM

Clean the MSM as follows:

1 Disconnecting the MSM from the IC system

Switch off the instrument.

Disconnect the MSM from the separation column, peristaltic
pump and detector.

2 Dismantling the MSM

Unscrew union nut (54-1) from the MSM housing (54-4).

Pull MSM connecting piece (54-2) and MSM rotor A (54-3) out of
the MSM housing (54-4). The MSM connecting piece and MSM
rotor A normally stick to one another - if this is not the case: Take
a sharp object, insert into the slot (54-5) in the MSM housing,
and pull out MSM rotor A (54-3) in this way.