2 cleaning the msm, Figure 54, Msm – components – Metrohm 850 Professional IC Anion MCS Prep 3 User Manual
Page 112
4.13 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
102
■■■■■■■■
850 Professional IC –Anion – MCS – Prep 3
4.13.3.2
Cleaning the MSM
It may be necessary to clean the MSM in the following cases:
■
Increased backpressure onto the output capillaries of the MSM.
■
Blockage of the MSM which cannot be eliminated (solutions can no
longer be pumped through the MSM).
■
Jamming of the MSM which cannot be eliminated (MSM can no longer
be switched over).
1
2
3
4
5
Figure 54
MSM – Components
1
Union nut
2
MSM connecting piece 6.2832.010
3
MSM rotor A 6.2832.000
4
MSM housing
5
Slot in the MSM housing
Cleaning the MSM
Clean the MSM as follows:
1 Disconnecting the MSM from the IC system
■
Switch off the instrument.
■
Disconnect the MSM from the separation column, peristaltic
pump and detector.
2 Dismantling the MSM
■
Unscrew union nut (54-1) from the MSM housing (54-4).
■
Pull MSM connecting piece (54-2) and MSM rotor A (54-3) out of
the MSM housing (54-4). The MSM connecting piece and MSM
rotor A normally stick to one another - if this is not the case: Take
a sharp object, insert into the slot (54-5) in the MSM housing,
and pull out MSM rotor A (54-3) in this way.