background image

■■■■■■■■■■■■■■■■■■■■■■

2 Installation

850 Professional IC –Anion – MCS – Prep 3

■■■■■■■■

41

1

3

2

3

4

5

6

7

8

2

1

Figure 23

SPM – Connectors

1

Union nut

2

SPM connecting piece (6.2835.010)

3

Inlet capillary
Made from PTFE. Labeled with Sample in.

4

Outlet capillary
Made from PTFE. Labeled with Sample out.

5

Rinsing solution inlet capillary
Made from PTFE. Labeled with H2O.

6

Rinsing solution outlet capillary
Made from PTFE. Labeled with Waste.

7

Regeneration solution outlet capillary
Made from PTFE. Labeled with Waste.

8

Regeneration solution inlet capillary
Made from PTFE. Labeled with Regenerant.

Caution

As the PTFE capillaries are very soft, the pressure screws should not be
overtightened.

Flattened capillaries should be shortened with the help of a capillary
cutter.