5.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
110
■■■■■■■■
850 Professional IC –Anion – MCS – Prep 3
5 Troubleshooting
5.1
Problems and their solutions
Problem
Cause
Remedy
Marked drop in
pressure
Leak in the system.
Check and seal connections (see Chapter 2.5,
page 13).
Marked rise in pres-
sure
6.2821.120 inline filter
blocked.
Replace 6.2821.130 filter (see Chapter 4.6,
page 87).
MSM – blocked.
■
Regenerate the MSM (see Chapter
4.13.3.1, page 100).
Note: 6.2821.180 pump tubing connection
with filter must be used (31-
).
Conductivity detector
blocked.
■
Shorten capillary ends by several mm (see
Chapter 4.15.1, Page 107).
■
Rinse detector against the normal flow
direction (see Chapter 4.15.1, Page 107).
Guard column – blocked.
Replace guard column (see Chapter 2.22,
page 66).
Separation column –
blocked.
■
Regenerate separation column (see Chap-
ter 4.16.4, page 108).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 69).
Note: Samples should always be microfiltered
(see Chapter 4.7, page 89).
Injection valve – valve
blocked.
Have the valve cleaned (by Metrohm service
technicians).
Drift of the baseline
Thermal equilibrium not yet
attained.
Condition (see Chapter 2.17, page 53) instru-
ment with the column thermostat switched
on .
Leak in the system.