Figure 24, Cation exchange with the spm, Figure 25 – Metrohm 850 Professional IC Anion MCS Prep 3 User Manual

Page 53: Injection and sequential suppression

background image

■■■■■■■■■■■■■■■■■■■■■■

2 Installation

850 Professional IC –Anion – MCS – Prep 3

■■■■■■■■

43

1

A

B

D

C

Figure 24

Cation exchange with the SPM

1

Sample

A

SPM

B

Injection valve

C

Separation column

D

MSM

2 Injection and sequential suppression

The injection valve is switched to the position Inject.

With the eluent flow (25-2) the sample prepared is rinsed onto
the separation column (25-C)
.

The sequential suppression with the MSM(25-D) and the MCS fol-
lows after the separation.

A

B

D

C

2

Figure 25

Injection and sequential suppression

2

Sample in the eluent flow

A

SPM

B

Injection valve

C

Separation column

D

MSM

Neutralization
A typical use of the SPM is the neutralization of an alkaline sample for
anion determination with chemical suppression.