beautypg.com

15 peristaltic pump, 1 principle of the peristaltic pump, Principle of the peristaltic pump – Metrohm 850 Professional IC Anion MCS Prep 3 User Manual

Page 56: Figure 29, Peristaltic pump, For suppression. see chapter 2.15, See chapter 2.15, For sample preparation. see chapter 2.15, F) and the mcs fol

background image

2.15 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

46

■■■■■■■■

850 Professional IC –Anion – MCS – Prep 3

D

Preconcentration column

E

Separation column

F

MSM

Note

In order to prevent impurities by the transfer water, a Metrosep I
Trap
(6.1014.200) can be installed between the injection valve 1 and
the peristaltic pump.

2.15

Peristaltic pump

2.15.1

Principle of the peristaltic pump

The Peristaltic pump is used for pumping sample and auxiliary solutions. It
can rotate in both directions.

The peristaltic pump pumps liquids according to the principle of displace-
ment. The pump tubing is clamped between the rollers (29-3) and the
tubing cartridge (29-5). During operation, the peristaltic pump drive
rotates the roller hub (29-2), so that the rollers (29-3) push the liquid for-
ward in the pump tubing.

6

1

4

3

2

5

7

Figure 29

Peristaltic pump

1

Knurled screw in the mounting pin

2

Roller hub

3

Rollers

4

Cartridge holder