beautypg.com

Sensor docking station, Analog sensor calibration – Bully Dog 4038x Sensor Docking Station User Manual

Page 29

background image

28

Sensor Docking Station

Analog Sensor Calibration

The Bully Dog Sensor Docking Station is capable of

supporting different brands of analog sensors. In order to

make sure that the information from your particular sensor is

being displayed correctly, you will need to change settings

in your GT or WatchDog based on the specifications of your

sensor. To calibrate the Sensor Docking Station for an analog

pressure sensor, you will need to know the

ratiometric output

.

Access the calibration settings through the following path on

your GT or WatchDog:

Main Menu

> User Options > SENSOR Settings

Set Input Low/High

This is the setting for the range of signal strength sent out

by the sensor. Manufacturers provide this information in the

specifications for each sensor; it is listed as ratiometric output,

and it must be between 0 and 5 volts.

Set Output Low/High

This is the setting for the scaling for the output for each

sensor: in other words, how the sensor’s electric signal

should be interpreted. This information is listed with the

ratiometric output in the specifications provided by the sensor

manufacturer, usually in parenthesis.

Example

You have installed an analog pressure sensor (oil, fuel or air) with the

following information listed in the manufacturer’s specifications:

0.5 to 4.5 volt ratiometric output

(0 psi at 0.5 volts and 100 psi at 4.5 volts,linear between these points)

You should enter the following information into the GT or WatchDog:
Set Input Low: 0.5 (volts)

Set Input High: 4.5 (volts)

Set Output Low: 0 (units)

Set Output High: 100 (units)