31

Appendix 3

User Applications and Wiring

This section is a collection of user specific applications and the wiring information required to activate them.

Connector Assignment

Table 1-20 BNC Jacks

Table 1-21 Terminal Strip

Connection

Description

Function

Levels

RF Power Input

Forward/Reflected

DC Input Forward &

Reflected

Detected RF signals

from sensors

30

μA full scale into

1400

Ω (42mV). May be

overranged to 100mV

Main Meter

Forward/Reflected

DC Main Meter Drive

Output

Current to drive

indicating meters

30mA full scale into

1400W (42mV)

Remote Meters

Forward/Reflected

DC Remote Meter

Drive Output

Voltage to drive

remote indicating

meters

1.038V full scale (use

33.2k resistor in

series with standard

Bird 30mA meters

Connection

Description

Function

Levels

TB1(+),

TB2(-)

External 12V Supply Used to allow mobile

operation

See Figure 18 on

page 35 for connection

information

TB3,4

Auxiliary Alarm Inputs Signals to trip alarms

from external sources

TTL levels, active low.

Pulled up to 5V by

4.7k internally.

Shunted to ground by

0.1

μF capacitor.

TB5

5V Output

Regulated 5V out put Can supply 360mA max.

TB6

Alarm Output

Low when alarm is

sounding, returned

high by reset

Active low. Pulled up to

5V by 4.7k. Will sink

1.5A and source 1mA.

TB7

Ground

Chassis ground

Reference for signals.

TB8

Reset Input /Output

Input to reset Wattcher

from external source.

May be used as an

output to reset other

equipment when

Wattcher is reset by the

front panel switch.

TTL levels, active low.

Pulled up to 5V by

4.7k internally.

Shunted to ground by

10

μF capacitor.

TB9

Confirm Output

Low when forward

power exceeds the

users confirm set point

Active low. Pulled up to

5V by 4.7k. Will sink

1.5A and source 1mA.

TB10

Active Forward
Monitor Input
(Carrier On)

Indicates to the

Wattcher that the RF

is on and that forward

power should be

monitored for a low

condition. There is an

adjustable internal

delay on the leading

edge of this signal.

TTL levels, active low.
Pulled up to 5V by
4.7k.