Record mode, Save the selected parameters – Rockwell Automation 1412-PP21xx PowerPad Portable Powermonitor User Manual
Page 42
42
Rockwell Automation Publication 1412-UM001D-EN-P - September 2012
Chapter 3
Display Modes
Record Mode
This mode enables all the parameters previously configured in the setup mode to
be recorded. Press the record display mode button .
See
for more information.
The variable function buttons have the following functions in this mode:
• Creates a new recording
• Opens a previous recording
• Deletes recording
Save the Selected Parameters
Recording status bar (displays remaining memory capacity).
Parameters - press the Left/Right Cursor buttons to select the parameters
and the Up/Down Cursor buttons to modify them. An underline will
appear under the selected parameter.
– Select the configuration number to be modified with the Up/Down
Cursor buttons. (CONFIG 1, 2, 3, or 4)
– Select the dates with the Up/Down Cursor buttons.
– Select a recording storage rate by using the Up/Down Cursor buttons.
– Enter the record name with the Up/Down Cursor buttons, which scroll
through the alphabet and numbers. Up to 7 characters may be entered.
– Press Enter to save the changes.
TIP
After a recording is set, the instrument will go into sleep mode (no display) to
save the batteries. The recording will start as programmed. Press any button
other than the On/Off button to turn On the display again.
TIP
The dates are adjusted according to the chosen recording integration
period. PERIOD does not refer to a sampling period, but rather, to an
integration period (average).
TIP
The possible storage rates are 1, 5, or 20 s; 1, 2, 5, 10, or 15 min.
TIP
The PowerPad unit calculates the storage needs of the recording, and if
necessary, displays the message ‘Not enough memory’.
07/25/02 10:56
100%
NEW RECORDING
CONFIGURATION : CONFIG 1
START :
07/25/02 10:56
END :
07/25/02 10:57
PERIOD :
1 min
NAME :
T
E
S T
Recording
2
1