Installation of the Q-Gard purification pack (Figure 4)
(The Q-Gard is only used on
Autopure WR600
systems equipped with a pack adapter)
1. Raise the pack adapter (A) to its highest position. Remove the two protective plastic inserts (B) on the
pack adapter.
2. Remove the two protective inserts on the Q-Gard pack (C). Wet the two O-rings with water.
3. Push the Q-Gard so that the pack adapter metal rod (D) goes through the hole at the top of the Q-Gard.
- Lift the Q-Gard slightly and push the bottom of it into the small opening (E) on the system.
- At the upper pack adapter, push the pack completely in until secure.
4. Lock the Q-Gard in place with the metal locking clip (F) clip on the end of the metal guide pin (D).
5. Bring the adapter cap down to its lowest position (G) so that it covers the top of the Q-Gard.
Electrical connection of system
1. Connect the power cord (Figure 2, B) to the Autopure WR600 (Figure 3,L). Connect the other end
of the power cord to an earth grounded power source.
2. Check that the point of use gun trigger is in the up-right position.
3. Turn on the system electrical switch (figure 1, H) by putting it to position I.
4. The serial number is displayed Note the serial in the table below.
Example:
Note : Contact Yamato Technical Service when connecting a level sensor to the Autopure WR600 system. As an option, it is
possible to get a “NO FEED WATER” message when there is no water in the reservoir.
Installation of the QUANTUM Polisher Cartridge (Figure 4)
1. Open the blue door on the front of the Autopure WR600 by pressing the 2 latches (H) on the door to
the left to open it.
2. Remove the two protective inserts from the QUANTUM cartridge. Wet the two rubber O-rings on the
QUANTUM polisher with water.
3. Install the QUANTUM polisher. Push it straight in so that the QUANTUM is fully inside.
4. Close the door. It is necessary to fully snap the latches (H) shut to hold the QUANTUM cartridge
inside.