Cleaning the A10

After performing several A-10 TOC measurements, the detection cell in the A-10 needs to be cleaned of any built- up residual
organic matter. If this build-up happens, then the displayed TOC values could be erratic or higher then previously seen.
To fix this, an autocleaning cycle is needed to oxidise any organic contaminants present in the A-10 detection cell.
To perform a cleaning cycle of the A-10, follow these instructions:



In PRE OPERATE mode

Press MENU for 2 seconds



Press MENU 3 times



Wait 5 seconds to validate the A-10 cycle. After 60
mins. the system will automatically return to its
initial operating mode.


(To interrupt the cleaning cycle)

Press MENU for 2 seconds,

to return to the initial operating mode.


Note: It is possible to get water from the point of use valve during the A-10 CLEANING mode

OPERATE/STANDBY MEASURE

CLEANING

MENU

P R O D . T I ME S E T U P

C O U N T E R :

0 m n

1 8 . 2 MΩ c m

T O C : 3 p p b

S E R V I C E : A 1 0

A 1 0 C L E A N I N G

6 0

OPERATE/STANDBY MEASURE

CLEANING

MENU

OPERATE/STANDBY MEASURE

CLEANING

MENU