beautypg.com

Appendix 1, Purging entrapped air from the uf cartridge, Interrupting the a-10 cleaning cycle – Yamato Scientific WR600S AutoPure User Manual

Page 26

background image

APPENDIX 1

Interrupting the sanitisation cycle of an UF cartridge


If a sanitant chemical has been introduced into the Autopure WR600 , it is absolutely necessary to let the sanitisation cycle finish.
If no sanitant has been introduced, then the sanitisation cycle can be cancelled by following these instructions.





Press CLEANING for 10 seconds

At the end of the sanitisation cycle, the will go back to its
initial mode of operation.


Purging entrapped air from the UF cartridge.


In either PRE OPERATE or PRODUCT

Press MENU for 2 seconds


press MENU 5 times




Open the point of use trigger to begin the air purge cycle.
The product water from the system will go to the drain via
the reject tubing.

At the end of the air purge, the system will go back into
PRODUCT mode.
Close the point of use trigger (vertical position).

Interrupting the A-10 cleaning cycle


If an A-10 cleaning cycle has been started from the MENU service mode, then it can be cancelled by pressing on the MENU
button.

P R O D . T I

M

E S E T U P

C O U N T E R :

0 m n

OPERATE/STANDBY MEASURE

CLEANING

MENU

S A N . C Y C L E 4 2 0 m n

S E R V I C E : U F

A I R P U R G E

OPERATE/STANDBY MEASURE

CLEANING

MENU

S T A N D B Y

OPERATE/STANDBY MEASURE

CLEANING

MENU