STATUS / PROBLEM
CAUSE
REMEDY
START SANIT.
A cleaning cycle of the UF cartridge
needs to be started.
Start a cleaning procedure.
See the MAINTENANCE chapter.
A10 ERROR #
A error has occurred with the A-10
TOC measuring equipment.
Press OPERATE / STANDBY to
reinitialise the Autopure WR600 . If the
error persists, then contact Yamato
Technical Service.
EXCHANGE A10 UV
The life duration of the UV lamp has
been reached.
Replace the UV lamp in the A-10.
Contact Yamato technical Service.
A10 CLEANING
The Autopure WR600 has the A-10
inside. The cartridges were just
replaced or an A-10 cleaning was
starting in the SERVICE MENU.
Let the Autopure WR600 finish the 60
minute A-10 cleaning cycle. The Autopure
WR600 can dispense water during this
mode.