Yamato Scientific WR600S AutoPure User Manual

Page 23

background image

STATUS / PROBLEM

CAUSE

REMEDY


START SANIT.


A cleaning cycle of the UF cartridge
needs to be started.


Start a cleaning procedure.
See the MAINTENANCE chapter.




A10 ERROR #



A error has occurred with the A-10
TOC measuring equipment.


Press OPERATE / STANDBY to
reinitialise the Autopure WR600 . If the
error persists, then contact Yamato
Technical Service.


EXCHANGE A10 UV


The life duration of the UV lamp has
been reached.


Replace the UV lamp in the A-10.
Contact Yamato technical Service.



A10 CLEANING


The Autopure WR600 has the A-10
inside. The cartridges were just
replaced or an A-10 cleaning was
starting in the SERVICE MENU.


Let the Autopure WR600 finish the 60
minute A-10 cleaning cycle. The Autopure
WR600 can dispense water during this
mode.