Connection of feed water to the system (figure 3), Connection of the reject tubing(s) (figure 3), Installation – Yamato Scientific WR600S AutoPure User Manual
Page 7
Installation
Connection of feed water to the system (Figure 3)
System fed from a reservoir :
1. Cut the feed water tubing, 8 mm OD (Figure 2, D) to the desired length (< 2.5 meters)
2. Remove the protective plug (B) from the “FEED 1” connection (H) by pressing on collar (A) and pulling
the plug out.
3. For wall mounting of the system, install elbow connectors (figure 2, J).
4. Connect the feed water tubing (C) (8 mm OD) to inlet “FEED 1”(H) by inserting it firmly in the fitting.
Verify that the connection is correctly made by pulling several times on the tubing once inserted.
5. Connect the other end of the feedwater tubing to the reservoir at an appropriate location ( Figure 5).
System fed from a pressurised source :
1. Cut the feed water tubing, 8 mm OD (Figure 2, D) to the desired length (< 2.5 meters)
2. Remove the protective plug (B) from the “FEED 1” connection (H) by pressing on collar (A) and pulling
the plug out.
3. The feed water valve or the fitting (D) should terminate in a 1/2” MNPT fitting.
The 1/2” FNPT fitting (E) is screwed on to the fitting (D). Use the Teflon tape (Figure 2, I), which is supplied with the
unit, to ensure a good seal against leaks.
4. Connect the tubing (C) to the fitting (F). Pull on the tubing afterwards to insure it is secure.
5. Pressurised feedwater entering the Milli-Q system must be regulated between 0.2 Bar (3 psi ) and 0.5 Bar
(7 psi) . Installation of a pressure regulator is necessary if the feedwater pressure is over 0.5 Bar (7 psi). The
regulator should be adjusted while water is dispensed from the system.
Connection of the reject tubing(s) (figure 3).
Any Autopure WR600 system with the A-10 TOC or UF option(s) has to have reject tubing on the system. The procedure
to connect the reject tubing is the same as that used for the feed water tubing.
1. Connect the Ultrafiltration cartridge reject tubing (8 mm OD) (Figure 2, D), < 2.5 m to the “DRAIN 3”
(I) outlet.
2. Connect the A10 waste stream tubing (6 mm OD) (Figure 2, E), < 2.5 m to the “OUT 5” (J) outlet.
Note: If the Autopure WR600 System is fed by a reservoir, the A10 reject water can be recycled.
When starting up the system, place all the ends of all reject tubing(s) to the drain.