Additional displays

System maintenance messages


DISPLAYED MESSAGE

ACTION SYSTEM

STATUS



EXCH. CARTRIDGES


Occurs automatically.
See MAINTENANCE chapter for
further information


SERVICE LED blinking
The operational lifetime of the purification /
polisher cartridges has expired.



START SANIT.


Occurs automatically.
See MAINTENANCE chapter for
further information


SERVICE LED blinking

A cleaning of the ultrafiltration module is
necessary.



EXCHANGE UV LAMP


Occurs automatically.
Call Yamato Technical Service


SERVICE LED blinking
The operational lifetime of the system UV lamp
has expired.



EXCHANGE A10 UV


Occurs automatically.
Call YamatoTechnical Service


SERVICE LED blinking

The operational lifetime of the UV lamp inside the
A10 has expired.



A10 CLEANING

59



Occurs automatically after installing
new purification / polisher pack(s). It
also can be started via the software.
See the MAINTENANCE chapter
for further information.



SERVICE LED blinking

A cleaning cycle of the A-10 is in progress.
Duration is 60 minutes.