Additional displays, System maintenance messages – Yamato Scientific WR600S AutoPure User Manual
Page 11
Additional displays
System maintenance messages
DISPLAYED MESSAGE
ACTION SYSTEM
STATUS
EXCH. CARTRIDGES
Occurs automatically.
See MAINTENANCE chapter for
further information
SERVICE LED blinking
The operational lifetime of the purification /
polisher cartridges has expired.
START SANIT.
Occurs automatically.
See MAINTENANCE chapter for
further information
SERVICE LED blinking
A cleaning of the ultrafiltration module is
necessary.
EXCHANGE UV LAMP
Occurs automatically.
Call Yamato Technical Service
SERVICE LED blinking
The operational lifetime of the system UV lamp
has expired.
EXCHANGE A10 UV
Occurs automatically.
Call YamatoTechnical Service
SERVICE LED blinking
The operational lifetime of the UV lamp inside the
A10 has expired.
A10 CLEANING
59
Occurs automatically after installing
new purification / polisher pack(s). It
also can be started via the software.
See the MAINTENANCE chapter
for further information.
SERVICE LED blinking
A cleaning cycle of the A-10 is in progress.
Duration is 60 minutes.