Yamato Scientific WR600S AutoPure User Manual
Page 19
This is a short sanitisation maintenance program. This allows the UF module to be sanitised at the start of an evening and
throughout that night.
Cycle 2:
This is a longer sanitisation maintenance program. This program allows for a more rigorous cleaning of the UF module whenever
flowrate is diminished from the system or the UF module is severely contaminated.
Before starting a sanitisation, it is necessary to have more than 25 litres of feedwater.
KEYPAD ACTION DISPLAY
Put the Autopure WR600 into STANDBY mode by
pressing OPERATE / STANDBY for 2 seconds
↓
Remove the MILLIPAK 40 from the point of use (figure 6)
↓
Screw the adapter fitting (figure 2, G) onto the point of use.
↓
Insert the 12 mm OD tubing (figure 2, F) onto the end of the barb fitting. Place the other end of this
tubing into a drain or sink.
↓
Remove the sanitisation port plug from the system. Keep it near the system
↓
Introduce 3 grams of sodium hydroxide into the sanitisation port.
↓
Screw the sanitisation port plug back onto the system. Check that it is secure.
↓
Press CLEANING for 2 seconds
(Pressing CLEANING again allows the choice of the longer
sanitisation cycle).
↓
Wait 10 seconds to validate the selection.
↓
Bring the point of use trigger forward to begin the
sanitisation cycle.
↓
At 400 minutes on the LCD, put the trigger
back to its upright position.
↓
At the end of the sanitisation cycle, the
system will automatically go into
PRE-OPERATE mode.
↓
Remove the tubing from the point of use. Remove the adapter fitting. Replace the MILLIPAK 40.
The Autopure WR600 system is now ready for normal use.