beautypg.com

Environmental conditions, Hydraulic specifications, Feedwater quality – Yamato Scientific WR600S AutoPure User Manual

Page 5: System performances

background image

Environmental Conditions


Ambient storage and operating temperature


Humidity


5°C < T < 40°C


20 % - 80 % without
condensation

Hydraulic specifications



Feedwater tubing

Reject tubing

Inlet feedwater pressure


Feedwater delivery flowrate

Feedwater temperature


8 mm outer diameter (OD), length : 3 meters maximum

8 mm and 6 mm OD, length : 2.5 meters maximum

Minimum: 0.1 bar (1.5 psi)
Maximum: 0.5 bar (7 psi)

≥ 1.5 litre/minute (LPM)

5 °C to 37 °C

Feedwater quality

Yamato recommends using water treated by distilled water or Reverse Osmosis.

System performances


Purified water quality


Resistivity 18.2

M

Ω-cm at 25 °C

Pyrogens

< 0.02 EU/ml

TOC*

<20

ppb (WR600A)

<5

ppb(WR600G)

<25

ppb(WR600B)

<10 ppb(WR600S)

Flowrates

Product water flowrate

Up to1.5 litre/minute

Up to 1.0 litre/minute

Sound level at 1 metre distance in dB

extremely low

l Test conditions: Autopure WR600 was equipped with a Q-Gard pre-treatment pack and a QUANTUM EX cartridge. The

feedwater to the Autopure WR600 came from a RiOs Reverse Osmosis system. TOC levels in the feedwater were < 50 ppb.
The quality of the Autopure WR600 product water can vary as a function of the quality of the feedwater.

This manual is related to the following products: