Yamato Scientific WR600S AutoPure User Manual

Page 25

background image

A-10

ERROR #

DESCRIPTION

REMEDY

0

EEPROM savings error


If the error persists, then contact Yamato Technical Service.

1

Error in analogue to digital

conversion


If the error persists, then contact Yamato Technical Service.

2

Temperature range error.

The temperature exceeded acceptable limits during analysis.
If the error persists, contact Yamato Technical Service.

3

The resistivity of the water in the

A-10 is not accepted at the

current temperature.

The correlation between these two measurements has passed the
allowed limits during the TOC measurement.
If the error persists, then contact Yamato Technical Service.

4

Temperature too cold.

The water temperature is below 5 °C. It is necessary that the
temperature be above this value.

5

Temperature too high.

The water temperature is above 41 °C. It is necessary that the
water temperature is below this temperature.

6

Conductivity too high


Conductivity of the sample water exceeded.

7

Overheating


The temperature exceeded 60 °C during oxidation.
If the error persists, then contact Yamato Technical Service.

8

Incomplete oxidation


The sample oxidation was not completed in an allotted time.
If the error persists, then contact Yamato Technical Service.

9

Low oxidation rate


The sample oxidation rate was abnormally low..
If the error persists, contact Yamato Technical Service.