Names and functions of parts – Yamato Scientific HF200 Heating Block User Manual
Page 7
![background image](/manuals/576631/7/background.png)
3
1. Names and functions of parts
Description of characters displayed in the temperature controller
The controller uses a display of 7-segment LED. Characters to display operation functions are
indicated in the “Measured Temp.” screen.
Character displays and corresponding meanings are described below.
Auto-stop or Auto-start operation can be selected with the TIMER key.
Character Identifier
Name
Meaning
AStP
Timer menu mode
Timer setting mode
display
Means that the auto-stop operation
has been selected.
End Time-up
Means that the auto-stop operation
has ended.
AStr
Timer menu mode
Timer setting mode
display
Means that the auto-start operation
has been selected.
Function settings as follows can be made from the sub-menu.
Keep pressing the TIMER key for four seconds, and the mode changes to the sub-menu mode.
To return to the original screen, keep pressing the TIMER key.
LocK
Sub-menu mode
Locked at fixed value
Function to lock so as to prevent
changing of the set contents
cAL
Sub-menu mode
Calibration offset
Temperature compensation function
to allow the aluminum block
temperature to agree with the
control temperature (see page 23)
Pon
Sub-menu mode
Power failure
compensation selection
Function to select whether the
recovery from power failure is to be
made automatically or manually.
on
Sub-menu mode
Locking the setting
Power failure
compensation selection
auto reset
Select ON with the ▲ key to
activate lock to prevent change of
the setting and when the recovery
from power failure is to be made
automatically.
off
Sub-menu mode
Canceling locking of the
setting
Power failure
compensation selection
manual reset
Select OFF with the ▼ key to
cancel locking to enable free
change of the setting and when the
recovery from power failure is to be
made manually.
Two characters to indicate that operation is not possible
nG Operation
impossible
Selection of the mode that is
inoperable during operation causes
display of nG in the set temperature
display, rejecting any attempt of
change.
Loc
Sub-menu mode
Locking the set contents
Attempt to change the mode while
the LocK function is ON causes
display of Loc in the set
temperature display, rejecting any
attempt of change.