2. How to operate the unit
Quick auto-stop operation
For quick auto-stop operation
In this mode, the fixed-value operation being under way is stopped automatically with the timer
setting.
t
Start with timer
Start of fixed-value operation Operation stop (automatic)
SV:Temperature set value t:Time
The quick auto-stop operation is performed in the fixed-value operation mode.
1. Start operation.
Measured Temp.
Fixed Temp.
Set Temp.
Auto Stop
Auto Start
Heater
Alarm
START
STOP
TIMER
SUB MENU 4sec.
℃
Press lightly the TIMER key during fixed-value
operation. The Set Temp. screen changes to display
the time and flashes, with the Auto Stop lamp entering
the flash state.
The time set immediately before the operation is
shown.
2.Set the desired time.
Measured Temp.
Fixed Temp.
Set Temp.
Auto Stop
Auto Start
Heater
Alarm
START
STOP
TIMER
SUB MENU 4sec.
℃
Set the desired time with ▼ and ▲ keys.
The timer can be set over a range from 0.01(1 minute) to 999.5(999 hours 50 minutes).
The timer indicates the time in minutes from 0.01 to 99.59 and in ten minutes from 100.0
to 999.5.
Timer setting
▼
SV