2 how to operate the unit, For fixed-value operation, Turn on the power switch to start power supply – Yamato Scientific HF200 Heating Block User Manual
Page 10: Set the temperature

6
2 How to operate the unit
Fixed-value operation(continuous operation)mode
For fixed-value operation
In this mode, the operation is started (by pressing the(START/STOP key for one second)and
continued at the set temperature till power supply is turned OFF (by pressing the START/STOP key
for one second).
SV:Temperature set value t:Time
1. Turn ON the POWER switch to start power supply.
Measured Temp.
POWER
Fixed Temp.
Set Temp.
Auto Stop
Auto Start
Heater
Alarm
OFF
START
STOP
TIMER
SUB MENU 4sec.
℃
The initial screen is shown for four seconds.
Measured Temp.
Fixed Temp.
Set Temp.
Auto Stop
Auto Start
Heater
Alarm
START
STOP
TIMER
SUB MENU 4sec.
℃
The temperature setting screen appears after the initial
screen has been shown for four seconds.
Shows the current bath temperature.
Shows the current set temperature.
2.Set the temperature.
Set the desired temperature with ▼ and ▲ keys.
The set temperature rises with the ▲ key and lowers
with the ▼ key.
Set the temperature over a range from the
room temperature + 5.0℃ up to 200.0℃.
Measured Temp.
Fixed Temp.
Set Temp.
Auto Stop
Auto Start
Heater
Alarm
START
STOP
TIMER
SUB MENU 4sec.
℃
▲
Power OFF
SV
▲
Operation start
t