6

2 How to operate the unit

Fixed-value operationcontinuous operationmode

For fixed-value operation

In this mode, the operation is started (by pressing the(START/STOP key for one second)and

continued at the set temperature till power supply is turned OFF (by pressing the START/STOP key

for one second).

SV:Temperature set value t:Time

1. Turn ON the POWER switch to start power supply.

Measured Temp.

POWER

Fixed Temp.

Set Temp.

Auto Stop

Auto Start

Heater

Alarm

OFF

START

STOP

TIMER

SUB MENU 4sec.

The initial screen is shown for four seconds.

Measured Temp.

Fixed Temp.

Set Temp.

Auto Stop

Auto Start

Heater

Alarm

START

STOP

TIMER

SUB MENU 4sec.

The temperature setting screen appears after the initial

screen has been shown for four seconds.

Shows the current bath temperature.

Shows the current set temperature.

2.Set the temperature.

Set the desired temperature with ▼ and ▲ keys.

The set temperature rises with the ▲ key and lowers

with the ▼ key.

Set the temperature over a range from the

room temperature + 5.0 up to 200.0.

Measured Temp.

Fixed Temp.

Set Temp.

Auto Stop

Auto Start

Heater

Alarm

START

STOP

TIMER

SUB MENU 4sec.

Power OFF

SV

Operation start

t