How to operate the unit – Yamato Scientific HF200 Heating Block User Manual

Page 25

background image

21

2. How to operate the unit

Other functions

Calibration offset function

The calibration offset function is to ensure agreement between the bath temperature and set

temperature by compensating the temperature difference, if any, between the bath temperature

actually measured with the reference thermometer or temperature recorder and the temperature set

with the controller (as read on the controller).

Enter the plus-side temperature difference to the calibration offset setting when the bath

temperature exceeds the set value and the minus-side temperature difference to the above setting

when the bath temperature is lower than the set level.

1.Measure the bath temperature

Wait till the bath temperature stabilizes, then check the difference between the bath temperature and

the measurement indicator of this device. The bath temperature can be measured correctly by using

the calibrated temperature recorder in the small amount of silicone oil filled into the aluminum block.

Current temperature

Temperature after compensation
on the minus side

Temperature after
compensation on the plus
side

Small amount (about 2 ml) of silicone oil

(for the aluminum block for 16.5 ㎜ test tube)

Temperature recorder sensor

Aluminum block