beautypg.com

How to operate the unit, Start operation, Stop operation – Yamato Scientific HF200 Heating Block User Manual

Page 11

background image

7

2. How to operate the unit

Fixed-value operationcontinuous operationmode

3. Start operation.

Measured Temp.

Fixed Temp.

Set Temp.

Auto Stop

Auto Start

Heater

Alarm

START

STOP

TIMER

SUB MENU 4sec.

Keep pressing the START/STOP key.

Fixed Temp. and Heater lamps go ON and the operation

begins.

When the bath temperature reaches the set

temperature, the Heater lamp flashes. Now the bath

temperature is under control.

With Fixed and Heater lamps ON, the bath temperature

begins rising and the operation begins.

4. Stop operation.

Measured Temp.

Fixed Temp.

Set Temp.

Auto Stop

Auto Start

Heater

Alarm

START

STOP

TIMER

SUB MENU 4sec.

Keep pressing the START/STOP key.

Fixed Temp. and Heater lamps go OFF and the

operation stops.

Fixed and Heater lamps go OFF and the operation

stops.

The temperature can be set over a range from 0.0 to 200.0.

Note however that the operating temperature range is from the room temperature + 5.0

up to 200.0.

Displays of the

operation panel

used for fixed-value

operation

POWER switch

▼ and ▲ keys

Measured Temp. screen

Set Temp. screen

Fixed Temp. lamp

Heater lamp

START/STOP key

With this switch ON, the power is supplied to

the system. With the switch OFF, power

supply is turned OFF. This switch is a

circuit breaker.

Key to change the set temperature.

Shows the current bath temperature.

Shows the set temperature.

Indicates that the fixed-value operation state

is under way.

ON with the heater energized and flashing

for the controlled state

Operation start/stop key. Keep pressing for a

certain period.

This manual is related to the following products: