10
2. How to operate the unit
Quick auto-stop operation
Displays of the
operation panel
used for quick
auto-stop operation
POWER switch
▼ and ▲ keys
Measured Temp. screen
Set Temp. screen
Fixed Temp. lamp
Auto Stop lamp
Heater lamp
START/STOP key
TIMER key
:
:
:
:
:
:
:
:
:
With this switch ON, the power is supplied to
the system. With the switch OFF, power
supply is turned OFF. This switch is a circuit
breaker.
Key to change the set temperature or the time
of timer
Indicates the current bath temperature.
Indicates the set temperature. Indicates the
set and remaining times of the timer.
Indicates that the operation is in the
fixed-value mode.
ON during auto-stop operation and flashing
for the time period set with the timer
ON with heater energized and flashing in the
controlled state
Operation start/stop key. Keep pressing for a
certain period.
This is also a timer start key for addition of
auto-stop timer operation in the fixed-value
operation mode.
Key to set the time during setting of the
auto-stop timer. Press the TIMER key lightly
to show the timer set time.