How to operate the unit, Other functions, Enter the offset temperature – Yamato Scientific HF200 Heating Block User Manual

Page 26

background image

22

2. How to operate the unit

Other functions

2.Select the calibration offset function from the sub-menu function.

Measured Temp.

Fixed Temp.

Set Temp.

Auto Stop

Auto Start

Heater

Alarm

START

STOP

TIMER

SUB MENU 4sec.

Keep pressing the TIMER key for four seconds to select

the sub-menu mode. First, the set-value lock function is

displayed. Press the TIMER key again, and

cAL

indicating the calibration offset function

appears.

The Set Temp. screen shows the compensation

temperature.

Shows cAL

indicating the calibration offset mode.

Shows the offset temperature set immediately before

operation.

3. Enter the offset temperature.

Enter the temperature difference on the minus side when the bath temperature is lower than the

temperature controller reading value.

Enter the temperature difference on the plus side when the bath temperature is higher than the

temperature controller reading value.

Repeat this step till the bath temperature and the temperature controller reading value agree.

Setting can be completely within a short period by first entering the compensation temperature

lower than the difference of both of them, then by allowing it to approach the target level gradually.

Measured Temp.

Fixed Temp.

Set Temp.

Auto Stop

Auto Start

Heater

Alarm

START

STOP

TIMER

SUB MENU 4sec.

Enter the temperature difference in the Set Temp.

screen.

Upon completion of the setting, keep pressing the

TIMER key. The screen returns to the temperature

setting screen.

Enter the temperature difference.