Operation method, Rs485 communication function – Yamato Scientific BB600 Low Constant Temperature Water Bath User Manual
Page 36
33
Operation Method
RS485 Communication Function
5. Communication Examples
5.1 Read communication example
Example) Request message:
A request for reading PV is transmitted to the unit set at address 02.
Reply message from the unit to this request message: The data of PV (00123) is returned.
S
T
X
0 2
R
P
V
1
E
T
X
B
C
C
Code Symbol/Data
ASCII
code
*2
①
Start Code
STX 02H
②
Address
02
30H 32H
③
Request (Read)
R 52H
④
Identifier
*1
PV1
50H 56H 31H
⑤
Numeric Data
00123
30H 30H 31H 32H 33H
⑥
End Code
ETX 03H
Request
61H
⑦
BCC data
Reply
02H
⑧
Acknowledgement code
ACK 06H
*1): See "7. List of Identifiers/Commands."
*2): For ASCII codes, see "8. List of ASCII Codes."
S
T
X
0
2
A
C
K
P
V
1
0
0
1
2
3
E
T
X
B
C
C
①
②
⑧
④
⑥
⑦
⑤
①
②
③
④
⑥
⑦
Read request message
(transmitted from the host computer)
Reply message (returned from the unit)