Operation method, Auto stop operation – Yamato Scientific BB600 Low Constant Temperature Water Bath User Manual
Page 23
![background image](/manuals/576607/23/background.png)
20
Operation Method
Auto Stop Operation
2. Start timer operation
• Press the RUN/STOP for one second after deciding the time.
• Timer operation starts with the AUTO STOP lamp lighting on.
• The timer is activated at the point when the temperature in bath
(measurement temperature) reaches to the setting temperature.
3. Stop/terminate timer operation
• The operation stops automatically at setting time.
• Buzzer continues to sound for about five minutes at operation
stop.
• The setting temperature screen displays the character "End",
which indicates termination of operation, with the FIXED TEMP
and AUTO STOP lamps lighting on. Press the RUN/STOP to
terminate the timer operation mode. The screen returns to the
initial setting screen.
To correct or check setting…
Changing the setting temperature or time during operation is possible by pressing the TIMER key. Use
the "▼▲" to change the setting value. Press the ENTER key respectively after changing the setting.
(Note that the time setting is required using the value calculated by adding a new additional time to the
already passed time in this case.)
Press the "▼" to display the setting temperature, operation mode and residual time on the setting
temperature screen.
When the dot is blinked, the indicator of the remaining time e.g."1.30" indicates the countdown. When the
dot is lit, the unit is under waiting (that is, the unit is under increasing or decreasing toward setting
temperature), and the timer stop s counting.