beautypg.com

Operation method, Auto start operation – Yamato Scientific BB600 Low Constant Temperature Water Bath User Manual

Page 24

background image

21

Operation Method

Auto Start Operation

In this mode, the unit automatically starts to operate after the set period passes away from the start
of fixed-value operation according to timer setting, as shown in the figure below. However, it does not

automatically come to a stop and must be manually deactivated.

温度(℃)

時間(t)

SV:温度設定値
PV:温度表示値

SV

スタート/ストップ キーON

スタート/ストップ キーOFF

タイマ起動

タイムアップ

PV

タイマスタート

Auto start operation
procedure

1. Set start time

Press the TIMER key on the initial screen.

The setting temperature display screen displays the character
"Astr", which indicates the auto start operation, with blinking.

Press the ENTER key.

The measurement temperature screen displays the character
"SV", which indicates the temperature setting. The setting
temperature screen displays the current setting temperature with
blinking. The AUTO START lamp blinks, too.

Set the temperature using the "▼▲".

The temperature can be set to the first decimal place.

Press the ENTER key again.

The measurement temperature display screen displays the
character "tim", which indicates the timer setting. The setting
temperature display screen displays the current setting time with
blinking.

Set the time using the "▼▲".

Timer function:

• The maximum setting time is "999 hours and 50 minutes".
• The time can be set in increments of a minute under 99 hours

and 59 minutes.

• It can be set in increment of ten minutes over 100 hours.
• The "▼▲"can change the setting time quickly when it is pressed

continuously. Press them discontinuously when fine adjustment
is needed.

TEMP

RUN/STOP key: ON

RUN/STOP key: OFF

TIME (t)

SV: Set Temp.
PV: Current Temp.

Timer starts

Time-up

Timer Operation

,④

,⑤