Operation Method

Calibration Offset Function

Calibration offset is a function which corrects the difference between the temperature in bath and
that of controller (sensor temperature) if arises. The function parallel corrects the difference either

to the plus or minus side within the whole temperature range of unit. The function can be set or

cancelled by the SUB MENU key.

Start operation with the target setting temperature. Check the

temperature in bath with a thermograph after it is stabilized.

Check the difference between the setting temperature and that in

bath.

Press the SUB MENU key. Select the character "cAL", which

indicates the calibration offset, using the "▲▼", and then press the
ENTER key.

Input the difference using the "▲▼" and then press the ENTER key.

This completes the setting.

™

The setting range of offset correction temperature is +99℃ to plus side

and -99℃ to minus side respectively.
When it is set to the minus side, the temperature on the measurement
temperature display screen falls by the setting temperature, while the
temperature on bath rises.
When it is set to the minus side, the temperature on the measurement
temperature display screen rises by the setting temperature, while the
temperature on bath falls.

™

The unit has two-point correction function, which performs offset

between low-temperature zone and high-temperature zone.

™

Please consult our local branch office when carrying out validation of

temperature controller.

Current temperature

Corrected temp. to plus side

Corrected temp. to minus side

,④

,④