beautypg.com

Operation method, Setting of overheating prevention device – Yamato Scientific BB600 Low Constant Temperature Water Bath User Manual

Page 18

background image

15

Operation Method

Setting of Overheating Prevention Device

The unit has the overheating prevention device (manual reset) that consists of independent
temperature measurement circuit, CPU, sensor and output circuit (it shares power source, display,

and key input with the controller) in addition to the automatic overheating prevention function

(auto reset) in the controller.

Setting range/function

The unit has failsafe functions against overheating. One of them is built in the controller and previously
set at factory shipment so to be automatically activated when the temperature exceeds the setting
temperature of temperature controller by 6℃, where the heater repeats on and off.
The other is united with the controller, which can be set by operating the keys on the controller.
The setting range of latter is from 0℃ to 50℃.
In case the temperature in bath exceeds the setting temperature of controller to reach to that of
overheating prevention device, the circuit is shut off and "Er19" is displayed with blinking on the screen of
controller with buzzer sound.
If the device is once activated,"Er19"continues to be displayed until the power is newly turned on.

Temperature setting procedure


1. Turn on the power (turn on the breaker in front)

• The default value is displayed for about four seconds after

turning on the power. The screen then displays the initial
setting. The current temperature in bath, operation mode
character and setting temperature of overheating
prevention device are displayed on respective screens.

2. Set the temperature for overheating prevention

Press the SUB MENU key.

Press the " ▼ ▲ " several times to select the setting

character of overheating prevention temperature "OH".

Press the ENTER key. The current setting temperature is

displayed with blinking on the setting temperature screen.

Note:

To prevent improper operation, set the value 10℃ or more

over the setting temperature of controller.

Select the value using the " ▼ ▲ "and then press the

ENTER key. This completes the setting.


Notes:

• The standard setting temperature of device is "the maximum setting temperature of unit plus

10℃" or "setting temperature plus 10℃". If the unit performs improper operation, increase it
5℃ more.

• The setting range of overheating prevention device is from 0℃ to 50℃. Improper setting of

temperature may cause inoperative of unit, malfunction of device, e.g. it is activated during
increasing in temperature in bath, or unexpected accidents such as fire disaster. To prevent
such matters, set a proper value.
The temperature is set to 90 at factory shipment.

• The purpose of overheating prevention device is to protect the unit from overheating. It does

not intend to protect the samples, or to protect them from the accident caused by the use of
explosive or inflammability.