Operation method, Rs485 communication function – Yamato Scientific BB600 Low Constant Temperature Water Bath User Manual
Page 31
28
Operation Method
RS485 Communication Function
2. Data Transmission Method
Item Specification
Communication standard
EIA standard, complying with RS-485
Synchronization method
Asynchronous communication method
Communication method
Half-duplex communication
Transmission code
ASCII code
Baud rate
1200/2400/4800/9600BPS
Communication distance
Max. 500 m (It depends on the effect of the ambient environment.)
Network
Multi-drop method (up 1:31 stations)
Signal wire
Two wires for transmission and receipt
Stop bit length
1/2bits
Data length
7/8bits
Parity
None/Odd/Even
BCC check
Enabled/Disabled
Response delay time
0 to 250msec
Communication address
1 to 99 stations (however, 1:31 stations at maximum)
Communication mode switching
RO/RW
Note) The shading indicates the initial setting of the unit.
3. Transmission Control Characters
Symbol Name Code
Detail
STX
Start of text
02H
Indicates the start of the text.
ETX
End of text
03H
Indicates the end of the text.
R
Read
52H
The command to read a request.
W
Write
57H
The command to write a request.
ACK
Acknowledge Character
06H
Transmits a reply when data is properly received.
NAK
Negative Acknowledge
15H
Transmits a replay in case of a receiving error.
Note)
R: Read (command to read settings or measured values)
W: Write (command to write set values)
R commands can be communicated at all times in all modes.
W commands can be communicated in regular mode only, and the parameters that can be set
depend on the operation state (during operation). See "7. List of Identifiers/Commands."
4. Transmission Control Procedures
4.1 Communication Procedure
・
This unit returns a "reply message" to a "request message" from the host computer but does not start
transmission.
・
This unit does not start communication (no reply) for about four seconds after the power is turned on.
Set a delay until communication begins after the power is turned on.