Operation method, Auto stop operation – Yamato Scientific BB600 Low Constant Temperature Water Bath User Manual
Page 22
19
Operation Method
Auto Stop Operation
In this mode, the unit automatically comes to a stop after the set period passes away from the start of
fixed-value operation according to timer setting, as shown in the figure below.
温度(℃)
時間(t)
SV:温度設定値
PV:温度表示値
SV
スタート/ストップ キーON
スタート/ストップ キーOFF
タイマ起動
タイマスタート
タイムアップ
PV
Auto stop operation
procedure
This operation is used to specify the automatic stop time in the
fixed temperature operation.
1. Set stop time
①
Press the TIMER key on the initial screen.
The setting temperature display screen displays the character
"AstP", which indicates the auto stop operation, with blinking.
②
Press the ENTER key.
The measurement temperature screen displays the character
"SV", which indicates the temperature setting. The setting
temperature screen displays the current setting temperature with
blinking. The AUTO STOP lamp blinks, too.
③
Set the temperature using the "▼▲".
The temperature can be set to the first decimal place.
④
Press the ENTER key again.
The measurement temperature display screen displays the
character "tim", which indicates the timer setting. The setting
temperature display screen displays the current setting time with
blinking.
⑤
Set the time using the "▼▲".
Timer function:
• The maximum setting time is "999 hours and 50 minutes".
• The time can be set in increments of a minute under 99 hours
and 59 minutes.
• It can be set in increment of ten minutes over 100 hours.
• The "▼▲"can change the setting time quickly when it is pressed
continuously. Press them discontinuously when fine adjustment
is needed.
TEMP
RUN/STOP key: ON
RUN/STOP key: OFF
TIME (t)
SV: Set Temp.
PV: Current Temp.
Timer starts
Time-up
Timer Operation