Operation method, Rs485 communication function – Yamato Scientific BB600 Low Constant Temperature Water Bath User Manual
Page 30
27
Operation Method
RS485 Communication Function
1. Settings Relating to Communication
1.1 Communication Settings
Before starting communication with the VS3 controller (hereinafter called the "unit"), set
communication parameters on the personal computer.
Item Communication
setting
1
Data length
8 bits
2
Stop bit length
2 bits
3 Parity
Disabled
4 BCC
check
Enabled
5 Baud
rate
4800BPS
6
Response delay time
0msec
1.2 Communication Connections
■
Personal computer
・
Use channel 1 (COM1/COM2 port) of the RS232C interface.
■
RS232C/RS485 converter
・
For the converter, System Sacom's KS-485 is recommended.
・
Our optional accessory "external communication adapter (RS485-232C) ODK18" permits the
connections described in Note 1) below (except the personal computer). A sample program is
uploaded on our website.
http://www.yamato-net.co.jp/support/program/index.htm
■
Communication cable for connection
PC
VS3 controller
CN3-3,4
RS232C/RS485 converter
KS-485
Cable 2
Cable 1
AC adapter
COM1/COM2
VS3 built-in device
Note)
The optional accessory "external communication adapter (RS485-232C) ODK18" comprises the
following:
①
Communication cable 1: One-meter-long RS-232C cable with a connector (for IBM nine-pin
appliance connection) to the personal computer and System Sacom's CBL16 connector (Dsub
25-pin male) to the KS-485
②
Communication cable 2: Three-meter-long UL2464TASB two-core AWG20 cable with a connector
(Dsub nine-pin male) to the KDS-485 and a Y-terminal (with a 100W terminating resistor) to the
unit
③
RS-232C <=> KS-485 conversion unit: System Sacom's KS-485 with an AC adapter