27

Operation Method

RS485 Communication Function

1. Settings Relating to Communication

1.1 Communication Settings

Before starting communication with the VS3 controller (hereinafter called the "unit"), set
communication parameters on the personal computer.

Item Communication

setting

1

Data length

8 bits

2

Stop bit length

2 bits

3 Parity

Disabled

4 BCC

check

Enabled

5 Baud

rate

4800BPS

6

Response delay time

0msec

1.2 Communication Connections

Personal computer

Use channel 1 (COM1/COM2 port) of the RS232C interface.

RS232C/RS485 converter

For the converter, System Sacom's KS-485 is recommended.

Our optional accessory "external communication adapter (RS485-232C) ODK18" permits the
connections described in Note 1) below (except the personal computer). A sample program is
uploaded on our website.
http://www.yamato-net.co.jp/support/program/index.htm

Communication cable for connection

PC

VS3 controller

CN3-3,4

RS232C/RS485 converter

KS-485

Cable 2

Cable 1

AC adapter

COM1/COM2

VS3 built-in device

Note)

The optional accessory "external communication adapter (RS485-232C) ODK18" comprises the
following:

Communication cable 1: One-meter-long RS-232C cable with a connector (for IBM nine-pin

appliance connection) to the personal computer and System Sacom's CBL16 connector (Dsub
25-pin male) to the KS-485

Communication cable 2: Three-meter-long UL2464TASB two-core AWG20 cable with a connector

(Dsub nine-pin male) to the KDS-485 and a Y-terminal (with a 100W terminating resistor) to the
unit