Zero the Transducer Interface

SWIFT

®

Mini TI

Installation

69

7. Push the Zero button on the front of the TI box. The upper Zero indicator

will turn on temporarily while the encoder angle is recorded. When the
lower Zero indicator turns on, proceed to the next step.

8. Rotate the tire in either direction until the lower Zero indicator turns off and

the upper indicator turns on (approximately 1.25 revolutions). The bridge
zeros will be computed using the average value of the bridges over one full
revolution. Once the computation is complete, both Zero indicators will
turn off.

Note

If the red Fail indicator lights a problem was detected during the zero
process. Try repeating the procedure. Use TI2STATUS for a more
detailed explanation of the problem. If you continue to have an error,
consult the chapter

“Troubleshooting”