Ti2status - transducer interface status, Description of ti2status indications, Ti2status - transducer interface status 31 – MTS WIFT Mini Transducer Interface User Manual
Page 31: Description of ti2status indications 31
TI2STATUS - Transducer Interface Status
SWIFT
®
Mini TI
Software Utilities
31
TI2STATUS - Transducer Interface Status
This program gets status information from the SWIFT Transducer Interface (TI)
when the TI has encountered a problem and the red failed indicator is lit. You can
use this program to easily interpret the error. For certain errors this program may
provide additional information.
Syntax
ti2status
The following is an example of the ti2status command report:
Example
Description of TI2STATUS Indications
Fatal Error:
Provides an indication that the CPU is unable to run. The message indicates the
possible reason (see
on page 38 for a list of the possible
errors).
Boot Loader version:
The boot loader is a program that verifies that the main program is complete. The
version number is a reference for use by service personnel (see
on page 38 for a list of the possible errors).
FPGA version:
Identifies the version of the functions programmed in the field programmable
gate array. The version number is a reference for use by service personnel.
Firmware Version:
Displays the version of the software installed in the TI. This software is stored in
flash memory and if needed can be upgraded in the field by an MTS service
engineer.
C:\bin>ti2status
SWIFT Mini TI status (Version 1.3)
Fatal error: NONE
Boot Loader version: 1
FPGA version: 10
Firmware version: 24
Zero: Good
Fx1 Shunt: Good
Fx2 Shunt: Good
Fy1 Shunt: Good
Fy2 Shunt: Good
Fy3 Shunt: Good
Fy4 Shunt: Good
Fz1 Shunt: Good
Fz2 Shunt: Good
Calibration data is: Good
Unit serial number: 02036413