beautypg.com

Ti2status - transducer interface status, Description of ti2status indications, Ti2status - transducer interface status 31 – MTS WIFT Mini Transducer Interface User Manual

Page 31: Description of ti2status indications 31

background image

TI2STATUS - Transducer Interface Status

SWIFT

®

Mini TI

Software Utilities

31

TI2STATUS - Transducer Interface Status

This program gets status information from the SWIFT Transducer Interface (TI)
when the TI has encountered a problem and the red failed indicator is lit. You can
use this program to easily interpret the error. For certain errors this program may
provide additional information.

Syntax

ti2status

The following is an example of the ti2status command report:

Example

Description of TI2STATUS Indications

Fatal Error:

Provides an indication that the CPU is unable to run. The message indicates the
possible reason (see

, “Error Messages,”

on page 38 for a list of the possible

errors).

Boot Loader version:

The boot loader is a program that verifies that the main program is complete. The
version number is a reference for use by service personnel (see

, “Error

Messages,”

on page 38 for a list of the possible errors).

FPGA version:

Identifies the version of the functions programmed in the field programmable
gate array. The version number is a reference for use by service personnel.

Firmware Version:

Displays the version of the software installed in the TI. This software is stored in
flash memory and if needed can be upgraded in the field by an MTS service
engineer.

C:\bin>ti2status

SWIFT Mini TI status (Version 1.3)

Fatal error: NONE
Boot Loader version: 1
FPGA version: 10
Firmware version: 24

Zero: Good

Fx1 Shunt: Good
Fx2 Shunt: Good
Fy1 Shunt: Good
Fy2 Shunt: Good
Fy3 Shunt: Good
Fy4 Shunt: Good
Fz1 Shunt: Good
Fz2 Shunt: Good

Calibration data is: Good

Unit serial number: 02036413