beautypg.com

Ti2xfer - transducer interface transfer, Ti2xfer - transducer interface transfer 33 – MTS WIFT Mini Transducer Interface User Manual

Page 33

background image

TI2XFER - Transducer Interface Transfer

SWIFT

®

Mini TI

Software Utilities

33

TI2XFER - Transducer Interface Transfer

This program is used to read the current settings in the TI and save them to the
computer (upload) or write the values from a calibration file on the computer to a
TI (download).

Syntax

ti2xfer

The following is an example of the ti2xfer command:

Example

File Format

The file used with commands contains a header and version number, and a list of
parameters. Transducer calibrations can be uploaded and saved individually. If a
test needs to be rerun at a later date but the original transducer is not available,
another transducer can be used by downloading its calibration information.

Following the header is a list of parameter settings. The syntax is:

ParamName=ParamValue

The following rules apply:

Tabs and spaces are allowed.

The parameters can occur in any order.

Names are case insensitive. If a parameter name is not recognized, a
warning will be reported.

If an error causes the program to abort while downloading, any parameters
prior to the error will have been successfully downloaded because
parameters are downloaded as they are read. Only those parameters in the
file are downloaded. Parameters not in the file are unchanged.

C:\bin>ti2xfer

SWIFT Mini TI transfer (Version 1.0)
Upload and download settings

0...Exit
1...Upload settings from TI box
2...Download settings to TI box

Choice? 1

Filename? sample.cal

0...Exit
1...Upload settings from TI box
2...Download settings to TI box

Choice? 0