Error messages, Error messages 38, Possible reason (see – MTS WIFT Mini Transducer Interface User Manual
Page 38: Error, Messages, Ge (see, 38 for a list of, Fer to
SWIFT
®
Mini TI
38
Error Messages
Software Utilities
Error Messages
When a SWIFT utility encounters an error, the red failed indicator on the TI front
panel lights. Run the TI2STATUS program to identify the cause of the error.
Following is a list of possible error messages:
•
Fatal Errors
NONE
Failed command port initialize
Failed I2C (integrated-integrated-circuit) initialize
Failed non-volatile memory initialize
Failed compute module initialize
Failed ADC (analog-to-digital converter) initialize
Failed DAC (digital-to-analog converter) initialize
Failed calibration multiplexer initialize
Failed while reading I2C
Failed timer initialize
Failed interrupt initialize
Failed encoder initialize
Failed zero module initialize
Failed shunt module initialize
Unrecognized failure code – This error could be due to hardware
failures or software bug.
•
Boot Loader – The boot loader runs when you turn on power. This line
displays the version number if boot was successful.
Error, boot loader has not run – This occurs if the CPU is not
running.
Error, the boot loader has not finished – Something prevented the
CPU from completing booting the main program.
Error, the boot loader found a bad Flash CRC (cyclic redundancy
check) – This error is likely due to a corrupted Flash program.
Computed CRC =
Flash CRC =
Error, unrecognized boot loader error code of: – Possibly corrupted
memory; communications or the version of TI2STATUS is not
compatible with the firmware.