Upload the calibration file, Upload the calibration file 47 – MTS WIFT Mini Transducer Interface User Manual
Page 47

Upload the Calibration File
SWIFT
®
Mini TI
Transducer Interface Setup
47
Upload the Calibration File
A unique calibration file was loaded into the TI memory by MTS before the
transducer and transducer interface were shipped. Use the program TI2XFER to
retrieve the calibration file.
1. Connect a USB cable from the laptop computer or PC to the TI.
Note
Ensure the proper USB 2.0 drivers are installed on the laptop or PC.
Refer to
on page 42, as necessary
2. Insert the SWIFT Transducer Interface Utilities CD or diskette into the
laptop computer or PC.
3. Run the program TI2XFER.
4. Enter 1 at the prompt to upload the calibration file. (See the illustration
below.)
5. Enter a file name.
6. TI2XFER will prompt you when the file has uploaded.
7. Enter 0 at the prompt to exit the program.
SWIFT Mini TI transfer (Version 1.0)
Upload and download settings
0...Exit
1...Upload settings from TI box
2...Download settings to TI box
Choice? 1
Filename? sample.cal
0...Exit
1...Upload settings from TI box
2...Download settings to TI box
Choice? 0