Zero the Transducer Interface
Installation
5. Adjust the adapter plate, as necessary, until the inclinometer reads 0.0°,
±0.1° (or 90.0°, ±0.1° depending on the orientation of the inclinometer) as
shown in the figure above.
6. Push the Zero button on the front of the TI box.
The Zero indicators will toggle during the zero process. When the Zero
indicators turn off the process is complete for this angle.
7. Remove the level bracket/inclinometer assembly.
8. Repeat Step 5 through 8 three more times.
Before Step 5 of each iteration, rotate the adapter plate 90°.
Always rotate the adapter in the same direction. (That is, if the first rotation
was clockwise, the succeeding rotations should also be clockwise.)
Note
If the red Fail indicator lights a problem was detected during the zero
process. Try repeating the procedure. Use TI2STATUS for a more
detailed explanation of the problem. If you continue to have an error,
consult the chapter
Setup for Horizontal Anti-rotate
Bracket Configuration