beautypg.com

Zeroing the transducer interface, For fixed mode, For spinning mode – MTS SWIFT 10 ATV Sensor User Manual

Page 92: Zeroing the transducer interface 92, Zeroing the transducer, Interface

background image

SWIFT 10 ATV Sensors

92

Road Simulator

Installing the Transducer

Zeroing the Transducer Interface

For fixed mode

For the non-spinning (fixed) zero method, use the TI2Xfer to download the fixed
calibration file (serialnumberf.ca)l to the appropriate TI box. The angle mode in
the file should be:

AngleMode=1

Rotate the transducer such that the orientation labeling is consistent with the
reference orientation. In most cases, this means rotating the transducer so the
labels are upright.

If an additional angle correction is required after installation, you will need to
measure the angle from zero, and then enter the new value for the AngleFixed in
the TI calibration file (see earlier instructions

, “Edit the Calibration File,”

on

page 57).

Zeroing the transducer in fixed mode can be performed before or after connecting
the adapter plate to the spindle housing.

Perform zero before connecting the adapter plate to the spindle housing if
you want to zero the transducer without the effects of the vehicle load on the
transducer.

With the transducer oriented as described above press Zero on the TI
Box.Repeat for each transducer.

Return to the procedure,

Attaching SWIFT Components to the Fixturing

to

complete the installation.

Perform zero after connecting the adapter plate to the spindle housing if you
want to apply the vehicle load to the transducer before you zero the
transducer.

Return to the procedure,

Attaching SWIFT Components to the Fixturing

to

complete the installation.

With installation complete and the vehicle load on the transducers, press the
Zero button on each TI Box.

For spinning mode

For the spinning zero method, use the TI2Xfer to download the spinning
calibration file (serialnumbers.cal) to the appropriate TI box. The angle mode in
the file should be:

AngleMode=0

This zero method samples all eight input bridges at four 90° intervals (that is at
0°, 90°, 180°, and 270°). After the data is taken, all eight input channels are
analyzed for signal offsets, and the X and Z input channels are analyzed to
determine the angular zero point.

1. Rotate the adapter plate with the transducer attached one full revolution so

that the encoder can find the zero index mark.