beautypg.com

Option 2, Option 3 – MTS SWIFT 10 ATV Sensor User Manual

Page 46

background image

SWIFT 10 ATV Sensors

46

TI2SHUNT - Transducer Interface Shunt

Software Utilities

Option 2

Use this to set the shunt tolerance. When selected, the following is displayed:

Option 3

Use this option to apply a shunt to each bridge individually, read the output of the
bridge, compare the result with the value stored in the Transducer Interface and
display the difference each bridge. See the table on the next page. The shaded
fields are the bridge being shunted. The actual sequence depends on the
transducer wiring. Note that this option does not update the shunt measured
values or error status.

Enter choice: 2

The current shunt tolerance is 2%

Enter new shunt tolerance in percent?