41

Fig. 42

2

1

3

5.2.4 Removal of the pump valves

The two lower pump valves (1) and (2) can be easily disassembled for
thorough drying. Valve (2) is located above valve (1).

For disassembly, support the valves (1) and (2) valve block at the pipe
end with an Allen key no. 3 at the screw (3) and unscrew the brass
valves with a screwdriver.

Dry the valves on a heating panel.

Before reinstallation, check whether the ball moves inside the valve
by shaking it.

5.