Cleaning and maintenance – Leica Biosystems LN22 User Manual

Page 38

background image

38

Instructions for Use V1.2, RevB – 10/2012

5.

Cleaning and maintenance

Use one of your hands to hold the cable loop
on the chamber bottom while inserting the
knife holder base together with the base plate
with your other hand. Pay careful attention
that the cable loop is not crushed between
bolts (21, Fig. 35) and the chamber bottom
plate.

Fig. 36

Fig. 37

9

17

17

11

17

12

15

Unscrew knife holder base (9) again, place on
edge in the chamber, fasten the chamber bot-
tom plate (11) with the screws (17).

Place eccentric disk (12) in the recess of the
bottom plate.

Loosely attach the chamber bottom plate (11)
to the knife holder base (9) with screws
(10, Fig. 31).

Reinsert the bottom plate (14) of the LN

2

trough

into the chamber using the manipulator.

Fig. 35

10

Insert the two screws with the washers in the
drillings (10) and screw the knife holder base
(9) onto the bottom plate.

Fig. 38

17

17