beautypg.com

2 maintenance, 1 resharpening the millers, Maintenance – Leica Biosystems SP2600 User Manual

Page 46: Cleaning, disinfection and maintenance

background image

46

Supplementary instruction manual V 1.2 – 07/2002

7. Cleaning, disinfection and maintenance

7.2 Maintenance

Only authorized service technicians may access the internal com-

ponents of the instrument for service and repair!

No liquids may enter the interior of the instrument during mainte-

nance!

• Carry out the maintenance steps to be done by the user as per the in-

structions in chapter 7.2, pages 67 - 70 of the Leica SM2500 instruction

manual.

• In addition to the maintenance steps listed in the Leica SM2500 manual,

the following maintenance work is required for the Leica SP2600:
chapter 7.2.1 - Resharpening the millers.
chapter 7.2.2 - Replacing the light bulb in the calibration device.

7.2.1 Resharpening the millers

• The millers need to be resharpened as soon as the milled specimen sur-

faces no longer have a smooth, mirrorlike appearance and/or when mill-

ing residues / residues of a grimy appearance start to accumulate on the

specimen surface.