2 maintenance, 1 resharpening the millers, Maintenance – Leica Biosystems SP2600 User Manual
Page 46: Cleaning, disinfection and maintenance
46
Supplementary instruction manual V 1.2 07/2002
7. Cleaning, disinfection and maintenance
7.2 Maintenance
Only authorized service technicians may access the internal com-
ponents of the instrument for service and repair!
No liquids may enter the interior of the instrument during mainte-
nance!
Carry out the maintenance steps to be done by the user as per the in-
structions in chapter 7.2, pages 67 - 70 of the Leica SM2500 instruction
manual.
In addition to the maintenance steps listed in the Leica SM2500 manual,
the following maintenance work is required for the Leica SP2600:
chapter 7.2.1 - Resharpening the millers.
chapter 7.2.2 - Replacing the light bulb in the calibration device.
7.2.1 Resharpening the millers
The millers need to be resharpened as soon as the milled specimen sur-
faces no longer have a smooth, mirrorlike appearance and/or when mill-
ing residues / residues of a grimy appearance start to accumulate on the
specimen surface.