beautypg.com

Operation – Leica Biosystems SP2600 User Manual

Page 41

background image

41

Leica SP 2600 – Ultramilling attachment

Recommended sledge speed settings for finishing milling
• Milling stroke: 0.5 mm/s to 1.0 mm/s

• Return stroke: any speed
Recommended rotational speed of miller for finishing milling
• < 2.000 rpm
Recommended mode of operation for finishing milling
• Select ‘continuous’ mode of operation.

Starting the milling procedure

During milling:

• switch on the vacuum cleaner,

• wear appropriate safety goggles,

• wear appropriate mask (type of mask depends on the material

being milled / size of the chips that form during milling).

Goggles and mask prevent milling chips and microdust from enter-

ing your eyes and from being inhaled.

• Switch on the vacuum cleaner (via button (35) on the control unit - see

chapter 5.4.11, page 34 of Leica SM2500 instruction manual) or via the

on/off switch of the vacuum cleaner itself.

• Start sledge stroke (buttons (29), (30) - chapter 5.4.10, p. 33 - Leica

SM2500 instruction manual).

• Carry out the milling steps as per above recommendation.

Important note:

If you need to mill several specimens with identical results (e.g.

for examining a specimen layer located at a certain level), com-

pletely identical parameters have to be selected for each speci-

men (for pre-milling as well as for the finishing milling procedure).

--> See following page for a perspective view of the finishing milling proce-

dure.

5. Operation

+