Operation – Leica Biosystems SP2600 User Manual
Page 39
39
Leica SP 2600 Ultramilling attachment
5. Operation
Ending the pre-milling procedure
Press RUN STOP or RUN ENABLE to stop the sledge (buttons (29), (30) -
Chapter 5.4.10, page 33 - Leica SM2500 manual) - or, when working in
PROG mode of operation, wait for the programmed number of strokes to
be finished (see chapter 5.4.3, page 26, Leica SM2500 manual).
If the specimen sledge has not stopped in the front limit position, activate
manual mode of operation MAN (button (5) - chapter 5.4.3, page 26,
Leica SM2500 manual) and move the specimen to the front limit position
(via buttons (58) or (59) - chapter 5.4.14, page 37- SM2500 manual) .
Stop miller rotation (button (37) - chapter 5.4.11, page 34 - Leica SM2500).
After switching off the milling spindle (button (37) - chapter. 5.4.11,
p. 34 - instruction manual Leica SM2500) the miller continues to
rotate for a little while!
Risk of injury!
Slide plexiglass protective cover to upper position.
Never touch the milling spindle while it is rotating and/or while
the plexiglass cover is shut, no matter whether the spindle is ro-
tating or not.
Remove the miller holder containing the pre-miller.
or