beautypg.com

Operation – Leica Biosystems SP2600 User Manual

Page 20

background image

20

Supplementary instruction manual V 1.2 – 07/2002

5.2 Adjusting the pre-millers in the

calibrating device

Preparatory steps
• Connect mains power supply (20) to miller cali-

bration device (18).

• Connect mains power supply (20) to mains: the

light bulb is switched on.

• Place magnifier (16) onto calibration device.
• Loosen regulating screws (6), (7) and (11) in

miller holder (2) until they are flush with the

inner surfaces of the miller holder.

If the screws protrude from the inner surfaces

of the miller holder (2), spring (1) and miller (3)

cannot be inserted into the correct position.

Twisted or deformed springs for the

miller holders must be exchanged!

Inserting the miller into the miller holder
• Insert spring (1) straight from above into miller

holder (2) - the two upper edges of the spring

must be flush with the upper surface of the

miller holder.

• Carefully insert miller (3) from the front - the

letter ‘V’ must be visible.

When the miller is inserted into the

holder the letter ‘V’ must be visible. -

That way, the milling edge is pointing

into the right direction.

• Insert the miller (3) completely - it must fit

against the rear and lower inner surface of the

miller holder (2) and be flush with spring (1).

• Insert screw (4) and washer (5) and tighten

slightly to fasten miller (3) .

Fig. 5.1

20

18

16

Fig. 5.2

6

6

11

7

1

3

2

5

4

5. Operation