Supplementary instruction manual V 1.2 – 07/2002

Adjustment along the y-axis

1 Select the same reference value as was set

for the pre-miller.

Pre-miller and finishing miller used for

milling a specific sample must be ad-

justed using the same reference value.

This ensures that both milling edges

hit the sample at exactly the same

height.

Minor differences in height can still be

adjusted via the height adjustment but-

ton on the control unit, once the miller

holder has been attached to the milling

spindle (see chapter 5.4.13, p. 36 the

Leica SM2500 manual).

2 To select the reference value, move lever (12),

setting the dial of the inner (smaller) scale (13)

to a position just below ‘1’ and the dial of the

outer scale (14) to approximately ‘70’ (approx.

10 o’clock) (Fig. 5.14a).

3 Rotate regulating screw (7) clockwise, bring-

ing the milling edge (3) close to the contact

plate (17).

4 Slowly continue to turn screw (7) clockwise

until the dial of scale (14) starts to move (i.e.

until the milling edge touches the contact

plate).

5 Keep rotating screw (7) very carefully and

slowly until the dial of scale (14) is in the up-

per ‘0’ position and the dial of scale (13) is lo-

cated exactly at ‘1’ (Fig. 5.14b).

Caution:

Be very careful when bringing the con-

tact plate close to the milling edge.

Always use the magnifier.

Make sure the milling edge does not

hit the contact plate!

12

14

13

17

3

7

14

13

Fig. 5.14 a

Fig. 5.14 b

5. Operation

7