6 preparatory steps prior to the milling procedure, Preparatory steps prior to the milling procedure, Operation – Leica Biosystems SP2600 User Manual
Page 34
34
Supplementary instruction manual V 1.2 07/2002
5.6 Preparatory steps prior to the milling procedure
For a detailed description of all control elements located on the
control unit, please refer to chapters 5.3 (Overview - control
panel) and 5.4 (Description of each individual control element) of
the Leica SM2500 instruction manual.
Select manual mode of operation MAN (button (5) - chapter 5.4.3, page
26, Leica SM2500 instruction manual).
In MAN mode, carry out the following preparatory steps:
Move the milling spindle to the upper limit position (button (48), chapter
5.4.13, p. 36, Leica SM2500 manual) and open the plexiglass cover.
Attached the miller holder (1) (containing the adjusted pre-miller) to the
milling spindle (use Allen screws size 5 (2) / Allen key size 5 - part of
SP2600 standard delivery).
Always slide plexiglass cover to upper position prior to attaching
the miller holders to the milling spindle!
5.
Operation
Fig. 5.22
1
2