Leica SP 2600 – Ultramilling attachment

• Insert the specimen (see chapters 5.5.2 and 5.5.4, pages 30 and 33 of this

manual) or - if you need to mill a vacuum table prior to first use - see

chapter 5.5.3, page 31 of this manual).

Wear protective gloves when working with potentially infectious

specimen material!

• Move the specimen sledge underneath the miller (buttons (56) or (57),

chapter 5.4.14, page 37, Leica SM2500 instruction manual).

• Set the milling window (button (61) and others - see chapter 5.4.14, page

37, Leica SM2500 instruction manual).

• Press button (51) (chapter 5.4.13, page 36, Leica SM2500 manual) to bring

the cutting edge of the miller close to the specimen surface (coarse ad-

justment - small gap must still remain).

Always slide the plexiglass cover to the upper position prior to the

distance check ‘milling edge / specimen surface’! - Risk of injury!

• Next, use buttons (50) and/or (49) (see chapter 5.4.13, page 36, Leica

SM2500 instruction manual) for fine adjustment of the distance between

milling edge and specimen surface.

• Check the remaining distance in three different zones of the specimen

surface (beginning/center/end) moving the specimen sledge via buttons

(56) - (59), see chapter 5.4.14, page 37, Leica SM2500 manual) - the cut-

ting edge of the miller should just about touch the specimen surface.

Caution:

Always check the distance between miller cutting edge and

specimen surface thoroughly prior to milling!

Under no circumstances may the miller cutting edge be positioned

more than a few microns (= selected milling thickness) below the

specimen surface when starting the milling procedure.

If the surface height of the specimen is uneven, the cutting edge

should just about touch the highest point of the specimen. - Then

mill the specimen until the surface is completely level.

or

5. Operation

or