KD Scientific Legato 200 Series User Manual
Page 78

78
Appendices
K
D
S
L
e
g
a
to
2
0
0
S
e
ri
e
s
U
s
e
r’s
M
a
n
u
a
l
KD Scientific Legato 200 Series Syringe Pump User’s Manual
Publication 5617-002-Rev F
Syringe Loading – Push-Pull Mechanism
phone
508.429.6809
fax
508.893.0160
web
www.kdscientific.com
Installing a Syringe
1.
Loosen Thumbscrews (1) to release the Retaining
Brackets (2 & 3) located on the Floating Syringe
Holder (4) and Pusher Block (5).
2. Loosen Syringe Clamp (6) by unscrewing the
Syringe Clamp Knob (7).
3. Place the syringe(s) under the Syringe Clamp on
the Floating Syringe Holder. Ensure that the
syringe’s flanges are within the Retaining Brackets.
4. Lock the flanges into place by tightening the
Retaining Brackets using the Thumbscrews.
5. Tighten the Syringe Clamp using the Syringe
Clamp Knob.
Repositioning the bracket location
1.
Unscrew the syringe clamp completely
2. Push the knob out of the clamp
3. Reposition and/or flip over and then push the knob
back into the clamp
1.
Thumbscrews
2.
Retaining Bracket,
barrel flange
3.
Retaining Bracket,
plunger flange
4.
Floating Syringe holder
5.
Pusher Block
6.
Syringe Clamp
7.
Syringe Clamp Knob
8.
Locking pins
9.
Stationary Syringe holder
1
4
8
9
6
7
2
3
5
Repositioning the Adjustable Syringe Holder
[For syringes > 60 ml]
For larger syringes you will need to detach the Floating
Syringe Holder and then mount it on the outside of the
Stationary Syringe Holder.
1.
Loosen the two Thumbscrews (1) on the Stationary
Syringe Holder (9) and slide the Retaining Bracket
(2) forward.
2. Remove the two Locking Pins (8)
3. Remove the Floating Syringe Holder (4).
4. Place the Floating Syringe Holder on the outside
of the Stationary Syringe Holder and lock into
place using the Locking Pins.
5. Slide the Retaining Bracket back and tighten using
the Thumbscrews.
Locking Pins